Helium Purging Enhances Trace Element Detection and Accuracy
in Soil Samples Using Bruker Tracer 5 Handheld X-Ray Fluorescence

X-ray fluorescence (XRF) is a nondestructive analytical technique used to determine elemental composition by measuring unique energy emissions patterns from atoms following inner-shell electron displacement. Due to its portability, handheld XRF devices are widely used in mining, manufacturing, and environmental monitoring. In addition, handheld XRF are increasingly being explored for forensic applications. However, their ability to detect light elements and achieve high sensitivity remains limited under ambient atmospheric conditions as air absorbs a significant portion of low energy X-rays.

This study investigates the optimization setup of helium purging to improve the detection sensitivity and accuracy of the Bruker Tracer 5 handheld XRF. To evaluate this, known standards previously analyzed using a XRF instrument from the Department of Soil and Crop Science were compared to the results of the handheld device under different variables such as mass. Results obtained with and without helium purging were compared to assess performance differences.

We found that introducing a helium environment will significantly enhance detection sensitivity and accuracy by reducing atmospheric interference with the X-Ray, particularly improving the detection of lighter elements such as F, Na, Mg, Al and Si. These findings could expand the applicability of handheld XRF in forensic science by enabling more reliable elemental analysis in field-based investigations..